前往化源商城

Analytical Sciences 2001-11-01

Flow-injection chemiluminometric determination of some thioxanthene derivatives in pharmaceutical formulations and biological fluids using the [Ru(dipy)3(2+)]-Ce(IV) system.

F A Aly, S A al-Tamimi, A A Alwarthan

文献索引:Anal. Sci. 17(11) , 1257-61, (2001)

全文:HTML全文

摘要

A flow-injection (FI) methodology using tris(2,2'-dipyridyl)ruthenium(II), [Ru(dipy)3(2+)], chemiluminescence (CL) was developed for the rapid and sensitive determination of three thioxanthene derivatives, namely zuclopenthixol hydrochloride, flupentixol hydrochloride and thiothixene. The method is based on the CL reaction of the studied thioxanthenes with [Ru(dipy)3(2+)] and Ce(IV) in a sulfuric acid medium. Under the optimum conditions, calibration graphs were obtained over the concentration ranges 0.002-6 migrograms/ml for zuclopenthixol hydrochloride, 0.5-15 micrograms/ml for flupentixol hydrochloride and 0.05-7.5 micrograms/ml for thiothixene. The limits of detection (s/n = 3) were 4.2 x 10(-9) mol/l zuclopenthixol hydrochloride, 2 x 10(-8) mol/l flupentixol hydrochloride and 4.5 x 10(-8) mol/l thiothixene. The method was successfully applied to the determination of these compounds in dosage forms and biological fluids.

相关化合物

结构式 名称/CAS号 全部文献
硫酸铯 结构式 硫酸铯
CAS:10294-54-9