Determination of cadmium in water samples by liquid electrode plasma atomic emission spectrometry after solid phase extraction using a mini cartridge packed with chelate resin immobilizing carboxymethylated pentaethylenehexamine.
Shigehiro Kagaya, Saori Nakada, Yoshinori Inoue, Waka Kamichatani, Hideyuki Yanai, Mitsuru Saito, Tamotsu Yamamoto, Yuzuru Takamura, Koji Tohda
Solid phase extraction using a mini cartridge packed with 22 mg of chelate resin immobilizing carboxymethylated pentaethylenehexamine was successfully utilized for separation/preconcentration of cadmium in water samples prior to liquid electrode plasma atomic emission spectrometric (LEP-AES) determination. The combined method with the extraction and LEP-AES was applicable to the determination of cadmium in the certified reference materials (EU-L-1 wastewater and ES-L-1 groundwater); the detection limit was 0.20 microg in 200 mL of sample solution (500-fold preconcentration).