Talanta 2005-01-01

Photometric detection in flow analysis systems using integrated PEDDs

Martina O’ Toole, King Tong Lau, Dermot Diamond, Martina O’ Toole, King Tong Lau, Dermot Diamond

文献索引:Talanta 66 , 1340-1344, (2005)

全文:HTML全文

摘要

A novel inexpensive optical-sensing technique has been developed for colorimetric flow analysis. This sensing system employs two LEDs whereby one is used as the light source and the other as a light detector. The LED used as light detector is reverse biased with a 5-V supply so that the photocurrent generated by the incident light discharges the capacitance. Direct digital output is provided by a simple timer circuit that measures the time taken for this discharge process from 5 V (logic 1) to 1.7 V (logic 0). This sensing concept has been applied in flow analysis by constructing an optical flow cell with a pair of LEDs. Calibration of the integrated optical flow cell using a dye resulted in a linear response that obeys the Beer–Lambert law. The flow rate, dynamic range, sensitivity and limits of detection were investigated. The system was also used for pH determination in the range of pH 2.5–6.8 using bromocresol green (BCG). The p K a of BCG was successfully determined by this technique.