P S Tofts, G J Barker, T L Dean, H Gallagher, A P Gregory, R N Clarke
Index: Magn. Reson. Imaging 15(1) , 69-75, (1997)
Full Text: HTML
The selection of a suitable low dielectric constant liquid for radiofrequency coil nonuniformity measurements is described. Measurements of dielectric constant (relative permittivity) were made on a range of candidate liquids. After excluding liquids that were too inflammable or too viscous, Esso Bayol 82 oil (dielectric constant epsilon' = 2.37) was chosen. At 1.5 T, a 27 cm diameter cylindrical phantom (test object) filled with Bayol 82 has a maximum nonuniformity of 1.9% arising from radiofrequency standing waves. The maximum diameter cylinder that can be used without the nonuniformity exceeding 2% is given for a range of liquids and field strengths. The construction of customized containers that fit closely inside a radiofrequency head coil from glass fiber reinforced resin ("fiberglass") is described. Thermal expansion of the liquid takes place without a rise in the internal pressure in the container.
Structure | Name/CAS No. | Molecular Formula | Articles |
---|---|---|---|
![]() |
petroleum ether
CAS:8032-32-4 |
C6H6 |
Antiinflammatory effect of petroleum ether extract of leaves...
1996-10-01 [J. Ethnopharmacol. 54(1) , 1-6, (1996)] |
The percutaneous absorption and skin distribution of lindane...
1997-11-01 [Hum. Exp. Toxicol. 16(11) , 652-7, (1997)] |
The percutaneous absorption and skin distribution of lindane...
1997-11-01 [Hum. Exp. Toxicol. 16(11) , 645-51, (1997)] |
White spirit--paint thinner, skin stripper.
2004-02-01 [Burns 30(1) , 86-7, (2004)] |
Larvicidal effects of mineral turpentine, low aromatic white...
2002-12-01 [Int. J. Dermatol. 41(12) , 877-80, (2002)] |
Home | MSDS/SDS Database Search | Journals | Product Classification | Biologically Active Compounds | Selling Leads | About Us | Disclaimer
Copyright © 2024 ChemSrc All Rights Reserved