Multispecies calibration: a novel application for inductively coupled plasma tandem mass spectrometry

Charles B. Williams, George L. Donati

Index: 10.1039/C8JA00034D

Full Text: HTML

Abstract

Inductively coupled plasma tandem mass spectrometry (ICP-MS/MS) has significantly expanded the reach of analytical atomic spectrometry. In this work, we take advantage of the gas-phase chemistry available in ICP-MS/MS to propose a novel method of calibration. Rather than employing several standard solutions and a single mass-to-charge ratio, multispecies calibration (MSC) uses only one standard reference concentration and several chemical species of a monoisotopic element for calibration. In this work, multiple oxide and ammonia species generated in an ICP-MS/MS collision/reaction cell are used to determine As, Co and Mn in rice and liver samples. Only two calibration solutions are required per sample: S1 is a 1 : 1 mixture of sample and standard solution; S2 has the same 1 : 1 volume ratio of sample and blank. They are run separately, and the analytical calibration curve is built by plotting signal intensities from several ions containing the analyte. S1 and S2 signals are plotted on the x-axis and y-axis, respectively, and each point in the calibration plot corresponds to a different analyte species. The instrumental limits of detection calculated for As, Co and Mn were 0.07, 0.03, and 0.07 μg L−1, with RSDs estimated as 7.8, 3.1 and 1.8%, respectively (n = 10). Certified and MSC-determined values of As, Co and Mn in Tomato Leaves (NIST 1573a) and Bovine Liver (NIST 1577b) presented no statistically significant differences (Student's t-test, 95% confidence level, n = 3). The MSC results were comparable and sometimes better than values determined by the traditional external standard, internal standard and standard additions calibration methods.

Latest Articles:

Use of an inline dilution method to eliminate species interconversion for LC-ICP-MS based applications: focus on arsenic in urine

2018-04-09

[10.1039/C8JA00038G]

Precise and accurate boron and lithium isotope determinations for small sample-size geological materials by MC-ICP-MS

2018-04-04

[10.1039/C7JA00400A]

A highly sensitive eco-scale method for mercury determination in water and food using photochemical vapor generation and miniaturized instrumentation for capacitively coupled plasma microtorch optical emission spectrometry

2018-04-03

[10.1039/C8JA00054A]

Antimony measurements in environmental matrices: seven considerations

2018-03-28

[10.1039/C7JA00391A]

An improved method of Cr purification for high precision measurement of Cr isotopes by double spike MC-ICP-MS

2018-03-28

[10.1039/C8JA00033F]

More Articles...