Acta Chimica Slovenica 2015-01-01

Simple and sensitive high performance liquid chromatography method with fluorescence detection for therapeutic drug monitoring of topiramate.

Daniela Milosheska, Tomaž Vovk, Iztok Grabnar, Robert Roškar

Index: Acta Chim. Slov. 62 , 411-9, (2015)

Full Text: HTML

Abstract

A simple, sensitive, accurate, precise and inexpensive HPLC method with fluorescence detection, suitable for routine therapeutic drug monitoring (TDM) of an antiepileptic drug topiramate, was developed and validated. The determination of plasma topiramate concentration was carried out after precolumn derivatization, using 4-chloro-7-nitrobenzofurazan as a fluorescent labeling agent and bendroflumethiazide as an internal standard. The standard calibration curve was linear over the concentration range of 0.01-24 μg/mL (r² > 0.9998). The intra- and inter-day accuracies expressed as bias were from 1.4 to 9.9% and from 1.9 to 10.2%, respectively. The intra- and inter-day precisions were below 7.9% and 2.7%, respectively. The validated method was applied for the measurement of plasma topiramate concentrations in patients with epilepsy. The reported method is appropriate for TDM of topiramate as well as for pharmacokinetic and bioequivalence studies.


Related Compounds

  • Sodium hydroxide
  • N-hexane
  • Acetonitrile
  • Methanol
  • Dichloromethane
  • Orthoboric acid
  • 3-Ethyl-2,4-pentan...
  • Formic Acid
  • ethyl acetate
  • 4-Chloro-7-nitrobe...

Related Articles:

Aptamer-based polyvalent ligands for regulated cell attachment on the hydrogel surface.

2015-04-13

[Biomacromolecules 16(4) , 1382-9, (2015)]

Osmoregulatory bicarbonate secretion exploits H(+)-sensitive haemoglobins to autoregulate intestinal O2 delivery in euryhaline teleosts.

2014-10-01

[J. Comp. Physiol. B, Biochem. Syst. Environ. Physiol. 184(7) , 865-76, (2014)]

Polymerization of affinity ligands on a surface for enhanced ligand display and cell binding.

2014-12-08

[Biomacromolecules 15(12) , 4561-9, (2014)]

Continuous syntheses of Pd@Pt and Cu@Ag core-shell nanoparticles using microwave-assisted core particle formation coupled with galvanic metal displacement.

2014-08-07

[Nanoscale 6(15) , 8720-5, (2014)]

Effect of (2)H and (18)O water isotopes in kinesin-1 gliding assay.

2014-01-01

[PeerJ 2 , e284, (2014)]

More Articles...