Determination of nine environmental phenols in urine by ultra-high-performance liquid chromatography-tandem mass spectrometry.
Minjian Chen, Pengfei Zhu, Bin Xu, Rencheng Zhao, Shanlei Qiao, Xiaojiao Chen, Rong Tang, Di Wu, Ling Song, Shoulin Wang, Yankai Xia, Xinru Wang
文献索引:J. Anal. Toxicol. 36(9) , 608-15, (2012)
全文:HTML全文
摘要
A method was developed to determine nine environmental phenols, including bisphenol A, 2,3,4-trichlorophenol, 2,4,5-trichlorophenol, pentachlorophenol, triclosan (2,4,4'-trichloro-2'-hydroxyphenylether), 4-tert-octylphenol, 4-n-octylphenol, 4-n-nolyphenol and benzophenone-3 (2-hydroxy-4-methoxybenzophenone) in human urine using ultra-high-performance liquid chromatography-tandem mass spectrometry (UPLC-MS-MS). The analytes were extracted and preconcentrated with solid-phase extraction, and then quantified with UPLC-electrospray ionization (negative ion mode)-MS-MS using multiple reaction monitoring mode. Limits of detection of the nine phenols ranged from 0.02 to 0.90 ng/mL. This method was further validated by the determination of phenols in 325 human urine samples that generated data regarding the exposure of various phenols to Chinese adults without occupational exposure to phenols.
相关化合物
相关文献:
2015-01-01
[PLoS ONE 10 , e0131428, (2015)]
2014-10-03
[J. Chromatogr. A. 1362 , 75-88, (2014)]
2015-01-15
[Sci. Total Environ. 503-504 , 69-86, (2015)]
2011-01-01
[Environ. Toxicol. Pharmacol. 31(1) , 198-204, (2011)]
2012-01-01
[J. Environ. Sci. (China) 24(9) , 1679-85, (2012)]