Direct aqueous injection LC-ESI/MS/MS analysis of water for 11 chloro- and thiomethyltriazines and metolachlor and its ethanesulfonic and oxanilic acid degradates.
Sung-Ben Huang, Thomas J Mayer, Robert A Yokley
文献索引:J. Agric. Food Chem. 56(8) , 2595-602, (2008)
全文:HTML全文
摘要
A multianalyte method is reported for the determination of atrazine, simazine, propazine, and their respective dealkylated chlorotriazine metabolites; ametryn and prometryn and their respective dealkylated thiomethyltriazine metabolites; and S-metolachlor and its ethanesulfonic and oxanilic acid degradates in deionized, ground, surface, and finished drinking water. Water samples are analyzed using direct aqueous injection (DAI) liquid chromatography-electrospray ionization/mass spectrometry/mass spectrometry (LC-ESI/MS/MS). No preanalysis sample manipulation is required other than transfer of a small portion of sample to an injection vial. The lower limit of the method validation is 0.050 microg/L (ppb) for all analytes except 2,4-diamino-6-chloro- s-triazine (didealkylatrazine, DDA, or G-28273). For this compound the LLMV is 0.50 microg/L (ppb). The overall mean procedural recoveries (and percent relative standard deviations) for all water types for all analytes ranged from 95 to 101% (4.5-11%). The method validation was conducted under U.S. EPA FIFRA Good Laboratory Practice Guidelines 40 CFR 160.
相关化合物
相关文献:
2015-01-09
[J. Chromatogr. A. 1376 , 35-45, (2015)]
2015-02-27
[J. Chromatogr. A. 1383 , 47-57, (2015)]
2016-01-15
[Sci. Total Environ. 541 , 33-41, (2015)]
2015-04-15
[Food Chem. 173 , 391-6, (2014)]
2015-11-11
[J. Agric. Food Chem. 63 , 9855-62, (2015)]