Improved accuracy in high-temperature conversion elemental analyzer δ18O measurements of nitrogen-rich organics.
Glendon B Hunsinger, Libby A Stern
文献索引:Rapid Commun. Mass Spectrom. 26(5) , 554-62, (2012)
全文:HTML全文
摘要
The use of high-temperature conversion (HTC) reduction systems interfaced with isotope ratio mass spectrometers for δ(18)O measurements of nitrogen-containing organic materials is complicated by isobaric interference from (14)N(16)O(+). This ion is produced in the ion source when N(2) reacts with trace oxygen shifting the m/z 30 baseline prior to elution of CO.We compared adaptations to a typical HTC system (TC/EA) to determine the best method to measure the δ(18)O values of nitrogen-rich organic substrates including: (1) 0.6 and 1.5 m 5 Å molecular sieve GC columns; (2) reduction of N(2) peak via He dilution; and (3) diversion of N(2) to waste via an automated four-port valve. These methods were applied to caffeine (IAEA-600), glycine, 4-nitroacetanilide, pentaerythritol tetranitrate (PETN) and cyclotrimethylene trinitramine (RDX), as well as pure and sodium azide-doped benzoic acid (IAEA-601) and sucrose (IAEA-CH6).The efficiency of N(2) production in the HTC interface was highly variable among these compounds. Both the longer column and the dilutor improved, but did not eliminate, the adverse effects of nitrogen.The diversion of N(2) adequately addressed the nitrogen-induced problems as indicated by: (1) consistent m/z 30 background offset between reference and sample CO for both N-free and N-rich materials; (2) production of the highest δ(18)O values; and (3) high correlation between the increase in the δ(18)O values relative to the GC-only measurements and the N(2) peak area. Additional validation would require N-rich oxygen isotope standards for inter-laboratory comparisons. Further, more stringent methodology may improve the poor inter-laboratory δ(18)O reproducibility of IAEA-600.Published in 2012 by John Wiley & Sons, Ltd.
相关化合物
相关文献:
2011-08-15
[Antioxid. Redox Signal. 15(4) , 925-31, (2011)]
2010-01-01
[J. Magn. Reson. 204(1) , 139-44, (2010)]
2011-01-07
[J. Chem. Phys. 134(1) , 014513, (2011)]
2012-12-01
[J. Colloid. Interface Sci. 387(1) , 175-9, (2012)]
2010-04-29
[J. Phys. Chem. A 114(16) , 5306-19, (2010)]