Avidin-EITC: an alternative to avidin-FITC in confocal scanning laser microscopy.
R Hulspas, P J Krijtenburg, J F Keij, J G Bauman
文献索引:J. Histochem. Cytochem. 41(8) , 1267-72, (1993)
全文:HTML全文
摘要
Detection of fluorescein-5-isothiocyanate (FITC)-labeled conjugates is suboptimal in two-color confocal scanning laser microscopy (CLSM). This limits the detection of small, dimly fluorescent targets. We explored the possible advantages of applying eosin-5-isothiocyanate (EITC) conjugated to avidin (Av-EITC) as an alternative for Av-FITC in CSLM. Despite the lower quantum efficiency of EITC, we found that the measured Av-EITC and Av-FITC emission intensities were similar as a result of the standard filter combinations used for simultaneous two-color detection in the Bio-Rad MRC 600 CSLM. The advantage of Av-EITC was that its fading characteristics compared very favorably to those of Av-FITC. An excitation intensity-dependent increase in Av-EITC fluorescence was observed, followed by an exponential decrease. This increase in fluorescence allows longer observation times, averaging of several scans without loss of brightness, and thus detection of dimly fluorescent targets by CSLM.