A stable and selective electrochemical biosensor for the liver enzyme alanine aminotransferase (ALT).
Mamun Jamal, Oliver Worsfold, Timothy McCormac, Eithne Dempsey
文献索引:Biosens. Bioelectron. 24(9) , 2926-30, (2009)
全文:HTML全文
摘要
An electrochemical method to determine alanine aminotransferase (ALT) activity over its normal and elevated physiological range was developed based upon detection of L-glutamate at a glutamate oxidase-modified platinum electrode. Measurements were carried out in the presence of ALT co-substrates L-alanine and alpha-ketoglutarate and current response from either the oxidation of hydrogen peroxide or the re-oxidation of the mediator ferrocene carboxylic acid was employed. The enzyme electrode was tested over a 6-month period and found to retain 79% of its original activity towards ALT detection with >200 measurements performed over this time. Signals associated with interfering electroactive species (ascorbic acid and uric acid) were eliminated using background subtraction at a denatured glutamate oxidase enzyme electrode. The sensitivity of the device was found to be 0.845 nA U(-1) L ALT with t(90)=180 s, linear range 10-1000 U L(-1) and LOD of 3.29 U L(-1) using amperometry at E(app)=0.4 V vs. Ag/AgCl at 308 K (35 degrees C).