Determination of HDL-cholesterol using 2,4,6-tribromo-3-hydroxybenzoic acid with a commercial CHOD-PAP reagent.
P Trinder, D Webster
文献索引:Ann. Clin. Biochem. 21 ( Pt 5) , 430-3, (1984)
全文:HTML全文
摘要
The Boehringer one-component high-performance cholesterol oxidase reagent has been modified by the inclusion of 2,4,6-tribromo-3-hydroxybenzoic acid (TBHBA) to give a fourfold increase in sensitivity to a molar absorbance of approximately equal to 29,000 with respect to cholesterol. The resulting reagent system is particularly suitable for the determination of plasma HDL-cholesterol for which a reagent of high sensitivity is required. A simple method of bromination avoiding the use of elemental bromine is used to prepare TBHBA. The modified reagent system has been found to have good within- and between-batch precision and has shown itself to be reliable and trouble-free.
相关化合物
相关文献:
1989-03-01
[J. Clin. Chem. Clin. Biochem. 27(3) , 151-6, (1989)]
1988-09-01
[Clin. Chem. 34(9) , 1799-804, (1988)]
1988-02-01
[Scand. J. Clin. Lab. Invest. 48(1) , 59-64, (1988)]
2010-01-01
[Microb. Cell Fact. 9 , 24, (2010)]
2015-04-01
[Am. J. Clin. Nutr. 101(4) , 725-32, (2015)]